| Catalog Number |
06P02-60 |
| Medical Device Problem Code |
False Positive Result (1227)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/05/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A follow up report will be submitted when the evaluation is complete.Section a1 patient identifier complete sid (b)(6).
|
| |
|
Event or Problem Description
|
|
The customer reported a false repeat reactive alinity s hbsag result on a tissue donor.The following results were provided: date sid alinity results nat result tissue disposition (b)(6) 2023, (b)(6), = 1.04 / 1.24 / 1.19 s/co, negative, discarded.Donor tissue discarded.
|
| |
|
Additional Manufacturer Narrative
|
|
The evaluation of complaint data for the product and likely cause alinity s hbsag reagent lot 49618fn00 identified normal complaint activity.Field data review showed that the repeat reactive rate observed for the complaint lot in blood banks meets the specificity performance claim of the package insert.Manufacturing documentation for the likely cause lot did not identify any issues associated with the complaint issue.No customer returns were available for evaluation.The performance of the likely cause lot was investigated by completing a review for non-conformances, potential non-conformances and deviations related to the likely cause lots.This review did not identify any non-conformances, potential non-conformances, or deviations.A review of labeling concluded that the issue is sufficiently addressed.No systemic issue and/or product deficiency was identified.
|
| |
|
Event or Problem Description
|
|
The customer reported a false repeat reactive alinity s hbsag result on a tissue donor.The following results were provided: date sid alinity results nat result tissue disposition 9/5/2023.(b)(6).= 1.04 / 1.24 / 1.19 s/co.Negative.Discarded donor.Tissue discarded.
|
| |
|
Search Alerts/Recalls
|