• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US UNIVERSAL STEM 75X18MM FLUTED; UNIVERSAL KNEE STEMS AND SLEEVES : KNEE STEMS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US UNIVERSAL STEM 75X18MM FLUTED; UNIVERSAL KNEE STEMS AND SLEEVES : KNEE STEMS Back to Search Results
Catalog Number 867418
Medical Device Problem Code Osseointegration Problem (3003)
Health Effect - Clinical Code Inadequate Osseointegration (2646)
Date of Event 10/10/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the pre-op diagnosis was aseptic loosening of the femoral component.The patient reportedly originally had a distal femoral fracture, followed by a primary tka, followed by a revision tka.The timeframes & further specifics were unknown.The tc3 femoral component along with the augments & stem extension were found to be loose.Loosening interface was bone/cement and cement manufacturer was unknown.Cultures run intra-operatively for infection were reportedly negative.The mbt rev tray w/sleeve & stem remained well-fixed & thus left in place.The femur was revised to an lps dfr w/a cemented stem along w/an lps art surface.Doi: unknown, dor: (b)(6) 2023, affected side: left knee.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.In order to determine if a lot related issue was possible, a worldwide complaint database search was performed.A worldwide complaint database search found no additional related reports against the provided product code/lot code combination.Based on the inability to find any additional related reports against the provided product code/lot code combination it is reasonable to conclude that there are no anomalies with regard to manufacturing or inspection contained in the device history records that would contribute to the reported event.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIVERSAL STEM 75X18MM FLUTED
Common Device Name
UNIVERSAL KNEE STEMS AND SLEEVES : KNEE STEMS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key17948810
Report Number1818910-2023-21111
Device Sequence Number7623935
Product Code JWH
UDI-Device Identifier10603295217848
UDI-Public10603295217848
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2024
Device Catalogue Number867418
Device Lot Number525626
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/04/2023
Supplement Date Received by Manufacturer10/25/2023
Initial Report FDA Received Date10/17/2023
Supplement Report FDA Received Date10/25/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/20/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PFC SIGMA FEM POST AUG SZ4 4MM; PFC SIGMA FEM POST AUG SZ4 8MM; PFC*SIGMA DI AUG12MMCO,SZ4LEFT; PFC*SIGMA DI AUG12MMCO,SZ4LEFT; PFC*SIGMA TC3 FEM LEFT SZ4; SIGMA FEM ADAPTER 5 DEGREE; TC3 RP TIBIAL INSERT S4,20.0; UNK KNEE TIBIAL SLEEVE MBT; UNK KNEE TIBIAL TRAY SIGMA (RP) MBT; UNKNOWN KNEE FEMORAL ADAPTOR BOLT
Outcome Attributed to Adverse Event Required Intervention;
Patient Age70 YR
Patient SexMale
-
-