| Catalog Number |
07027770190 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/28/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The cobas e 801 analytical unit serial number was (b)(6).The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
There was an allegation of questionable rubella igg elecsys results from the cobas e 801 analytical unit.The initial result was 71.8 mui/ml (positive).The result ion another laboratory using abbott was negative.The sample was repeated on two cobas analyzer and the results were 69.8 mui/ml and 64.3 mui/ml (positive).No information was provided to determine if the questionable result was reported outside of the laboratory.
|
| |
|
Additional Manufacturer Narrative
|
|
The unit of measure was actually iu/ml.The investigation did not identify a product problem.The cause of the event could not be determined.
|
| |
|