DEPUY ORTHOPAEDICS INC US HP EM TIBIAL JIG SPIKED UPROD; KNEE INSTRUMENT : ALIGNMENT DEVICES Back to Search Results
Catalog Number 950501230
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/08/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that the front tibial spike broke off the device.
 
Additional Manufacturer Narrative
Product complaint (b)(4).Investigation summary according to the information received, "front tibial spike has broken off".¿ ¿the product was not returned to depuy synthes, however photos were provided for review.See attachment ¿(b)(4).Fw sigma base femur tibia (hp em tibia jig spiked-950501230)¿.¿ the photo investigation revealed that the long spike of the 950501230, hp em tibial jig spiked uprod has broken off.¿ the observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.¿ since the device was not returned, a dimensional inspection cannot be performed.¿ the overall complaint was confirmed as the observed condition of the 950501230, hp em tibial jig spiked uprod would contribute to the complained device issue.¿ based on the investigation findings, potential cause can be traced to unintended user error and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot :the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HP EM TIBIAL JIG SPIKED UPROD
Common Device Name
KNEE INSTRUMENT : ALIGNMENT DEVICES
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key17949500
Report Number1818910-2023-21131
Device Sequence Number6799418
Product Code HTZ
UDI-Device Identifier10603295225355
UDI-Public10603295225355
Combination Product (Y/N)N
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number950501230
Device Lot NumberSO2030952
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/09/2023
Supplement Date Received by Manufacturer10/18/2023
Initial Report FDA Received Date10/17/2023
Supplement Report FDA Received Date10/18/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/23/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-