• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. HALOGEN LIGHT SOURCE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. HALOGEN LIGHT SOURCE Back to Search Results
Model Number CLK-4
Medical Device Problem Code No Device Output (1435)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/27/2023
Type of Reportable Event Malfunction
Event or Problem Description
The olympus field service engineer (fse) on behalf of the customer reported to olympus, the halogen light source had no power.The issue was found during maintenance.There were no reports of patient harm.
 
Additional Manufacturer Narrative
The device was returned to olympus for evaluation and the customer's allegation was confirmed.The device evaluation found that the device was not turning on due to a faulty thermal switch, and that the lamp was not turning on due to a faulty transformer.Additional findings include the following: the device had a low air flow due to a faulty pump unit, heavy corrosion was found on all the metal parts (these parts could not be detached and needed to be replaced), third party repairs were evident on the lamp harness, the front panel was found to be broken, and heavy corrosion was found on the real panel / top cover / chassis.The investigation is ongoing and follow up with the user facility is currently being performed.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely that the event occurred due to a faulty thermal switch and a faulty transformer.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HALOGEN LIGHT SOURCE
Common Device Name
HALOGEN LIGHT SOURCE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17950434
Report Number3002808148-2023-11381
Device Sequence Number7530430
Product Code FAJ
UDI-Device Identifier04953170024511
UDI-Public04953170024511
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K993041
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCLK-4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/03/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/27/2023
Supplement Date Received by Manufacturer12/08/2023
Initial Report FDA Received Date10/17/2023
Supplement Report FDA Received Date12/12/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/13/1997
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-