Model Number PM3242
Medical Device Problem Code Loose or Intermittent Connection (1371)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/29/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that patient presented for routine generator change procedure.During procedure it was found that patient's atrial lead and right ventricular lead was not able to disengage from device header it was further noted that set screws of atrial and right ventricular port were stripped.The device was remained implanted.There were no patient consequences.
 
Event or Problem Description
New information received notes that the device was explanted and replaced.The patient was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLURE QUADRA RF CRT-P
Common Device Name
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key17951315
Report Number2017865-2023-49988
Device Sequence Number6799951
Product Code NIK
UDI-Device Identifier05414734507912
UDI-Public05414734507912
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P030035
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2016
Device Model NumberPM3242
Device Lot NumberA000002527
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/29/2023
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date10/17/2023
Supplement Report FDA Received Date11/09/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/28/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
QUARTET.; TENDRIL.; TENDRIL.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age85 YR
Patient SexFemale
-
-