| Model Number |
360-2080-01 |
| Medical Device Problem Code |
Failure to Obtain Sample (2533)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/29/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Attempt at intrahepatic biopsy which was a failure, because biopince did not recover liver material.
|
| |
|
Additional Manufacturer Narrative
|
|
The sample is indicated as available for return.As of the date of this report, the sample has not been returned.A follow-up report will be provided once the device has been received and reviewed.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the manufacturing and inspection records for this lot was conducted.No deviations or non-conformances were found.One opened sample was returned from the customer for review.A visual inspection was performed on the returned product, and it was found that the pincer was flattened, inhibiting it from properly obtaining a sample.The complaint was confirmed.There are stringent computer and photographic inspections that ensure the pincer is not deformed during the manufacturing process and has the proper radius.When the device is cocked, the pincer comes out of its window and becomes exposed to the tissues and structures present during the operation.It is also possible that the needle set was inserted into tissue that was too hard for the sensitive pincer, pushing it flat.Since the most likely cause for the damaged needles is related to an event within the user's environment and not a manufacturing issue, no corrective action can be implemented at this time.
|
| |
|
Event or Problem Description
|
|
Attempt at intrahepatic biopsy which was a failure, because biopince did not recover liver material.
|
| |
|
Additional Manufacturer Narrative
|
|
Udi related data quality updates only.
|
| |
|