| Catalog Number |
11131-000001 |
| Medical Device Problem Code |
Key or Button Unresponsive/not Working (4063)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/25/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned to stryker for evaluation.The cause of the reported issue could not be determined.
|
| |
|
Event or Problem Description
|
|
The customer contacted stryker to report that the shock button on their internal paddles handle was inoperable.The customer advised that during testing, the shock button did not work.In this state defibrillation therapy would be delayed or unavailable if needed.There was no patient use associated with the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
Stryker evaluated the customer's returned internal paddles and verified the reported issue.Stryker observed molding inside the cavity of the shock switch which impedes the movement of the button making it difficult to activate the shock button.The returned item was archived by stryker and the customer received a replacement.
|
| |
|
Event or Problem Description
|
|
The customer contacted stryker to report that the shock button on their internal paddles handle was inoperable.The customer advised that during testing, the shock button did not work.In this state defibrillation therapy would be delayed or unavailable if needed.There was no patient use associated with the reported event.
|
| |
|