H3: code "other" was selected as the medical device remains implanted.Return not possible.H6: code b14: a review of the manufacturing records for the device verified that the lot met all pre-release specifications.H6: code d12: according to the gore® excluder® iliac branch endoprosthesis instructions for use (ifu), adverse events that may occur and/or require intervention include, but are not limited to: endoleak.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
On (b)(6) 2019, this patient underwent an endovascular treatment for a left common iliac artery aneurysm using gore® excluder® iliac branch endoprosthesis (ibe).The patient¿s aorta was not suitable for stent graft treatment; thus, the physician decided to implant the ibe alone.An aortic extender component of gore® excluder® aaa endoprosthesis was added proximal to the ibe device in order to extend the sealing zone.No endoleak was found.The patient tolerated the procedure.On (b)(6) 2023, a reintervention was performed due to a type ib endoleak from the left internal iliac artery and common iliac aneurysm expansion.The left internal iliac artery was coil-embolized.A contralateral leg component (plc271000j) was additionally implanted upside down to reline the iliac branch component.The final angiography showed no endoleak.The patient tolerated the procedure.No opinion regarding the cause of endoleak was obtained from the physician.
|