| Model Number |
MI1210 SYNCHRONY ST |
| Medical Device Problem Code |
Impedance Problem (2950)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
09/14/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On (b)(6) 2023, the user came to the clinic for a routine appointment where channels with abnormal impedances were discovered.In addition, a good ecap response couldn't be observed.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: according to the currently available information, damage to the active electrode likely caused by minute device mobility is suspected.However, to determine an exact root cause a device investigation of the explanted device is necessary.No information on possible further steps has been received.
|
| |
|
Event or Problem Description
|
|
On 21-sep-2023, the user came to the clinic for a routine appointment where channels with abnormal impedances were discovered.In addition, a good ecap response couldn't be observed.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: according to the currently available information, damage to the active electrode likely caused by minute device mobility is suspected.However, to determine an exact root cause a device investigation of the explanted device is necessary.The concerned device was explanted but has not been received for investigation yet.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2023, the user came to the clinic for a routine appointment where channels with abnormal impedances were discovered.In addition, a good ecap response couldn't be observed.The user was re-implanted on (b)(6) 2024.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: damage to the active electrode which is consistent with minute device mobility was determined to have led to device failure over time due to fatigue wire breakages.The problems given in the recipient report appear to match the damage found.This is a final report.
|
| |
|
Event or Problem Description
|
|
On 21-sep-2023, the user came to the clinic for a routine appointment where channels with abnormal impedances were discovered.The user was re-implanted on (b)(6) 2024.
|
| |
|