| Model Number |
71004 |
| Medical Device Problem Code |
Improper or Incorrect Procedure or Method (2017)
|
| Health Effect - Clinical Code |
Inadequate Pain Relief (2388)
|
| Date of Event |
09/19/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
There are no allegations of any system or device failure or malfunction.Communication issues were reported immediately upon implanting the system, indicating that the ipg was implanted beyond the recommended depth.All implanted components remain in use and the patient reports being happy with the therapy after moving the ipg to a more superficial location.
|
| |
|
Event or Problem Description
|
|
Patient was implanted with the nalu peripheral nerve stimulator system on (b)(6) 2023 and immediately reported issues with communication between the implantable pulse generator (ipg) and the external therapy discs, leading to inadequate pain relief due to intermittent therapy, troubleshooting was attempted and unsuccessful.A surgical revision was performed to move the existing ipg to a more superficial location on (b)(6) 2023.The system was activated on (b)(6) 2023 with good connectivity and providing consistent therapy.
|
| |
|