• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDACTA INTERNATIONAL SA REVERSE SHOULDER SYSTEM GLENOID BASEPLATE Ø22X15; SHOULDER REVERSE SYSTEM GLENOID BASEPLATE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDACTA INTERNATIONAL SA REVERSE SHOULDER SYSTEM GLENOID BASEPLATE Ø22X15; SHOULDER REVERSE SYSTEM GLENOID BASEPLATE Back to Search Results
Catalog Number 04.01.0148
Medical Device Problem Code Positioning Failure (1158)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/12/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Batch review performed on 29 sept 2023: lot 2105497: (b)(4) items manufactured and released on 27-jul-2021.Expiration date: 2026-jul-08.No anomalies found related to the problem.To date, 12 items of the same lot have been sold with no similar reported event during the period of review.Additional involved implants: batch review performed on 29 sept 2023 on reverse shoulder system 04.01.0167 glenosphere 32xø22 (k170452) lot.179964b lot 179964b: (b)(4) items manufactured and released on 13-jan-2022.Expiration date: 2027-jan-11.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported event during the period of review.No anomalies have been found related to the problem on the lot of the semi-finished (glenosphere screw.) batch review performed on 29 sept 2023 on reverse shoulder system 04.01.0168 glenosphere 36xø22 (k170452) lot.2109111 lot 2109111: (b)(4) items manufactured and released on 11-oct-2021.Expiration date: 20266-sept-27.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported event during the period of review.No anomalies have been found related to the problem on the lot of the semi-finished (glenosphere screw.) batch review performed on 29 sept 2023 on reverse shoulder system 04.01.0177 glenosphere screw (k170452) lot.2215683 lot 2215683: (b)(4) items manufactured and released on 06-jul-2022.Expiration date: 2027-jun-19.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported event during the period of review.
 
Event or Problem Description
During the primary shoulder surgery, it was observed that the surgeon was not able to set the glenoid onto the baseplate twice.The locking screw would not screw into glenoid baseplate ø22x15; a new locking screw was opened because the one with the glenosphere was damaged but the coupling was still not possible.The surgeon changed the baseplate to a threaded glenoid baseplate ø24.5x25 and again the locking screw would not engage the baseplate.The surgeon switched to a lat.Glenosphere 36xø24.5, changed the poly and the surgery was completed successfully.There was a 2-hour delay.A consignment set was utilized.Additional anesthesia was administered.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REVERSE SHOULDER SYSTEM GLENOID BASEPLATE Ø22X15
Common Device Name
SHOULDER REVERSE SYSTEM GLENOID BASEPLATE
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, switzerland 6874
SZ  6874
Manufacturer (Section G)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, switzerland 6874
SZ   6874
Manufacturer Contact
marco giannessi
strada regina
castel san pietro, switzerland 6874
SZ   6874
MDR Report Key17957400
Report Number3005180920-2023-00806
Device Sequence Number17401811
Product Code PHX
UDI-Device Identifier07630040706353
UDI-Public07630040706353
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K170452
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 10/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/08/2026
Device Catalogue Number04.01.0148
Device Lot Number2105497
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/18/2023
Initial Report FDA Received Date10/18/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/27/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-