| Model Number |
32-2816 |
| Medical Device Problem Code |
Flaked (1246)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/07/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The user facility reported that during a patient procedure, a metal flake from their snowden-pencer diamond-tip wire/pin cutter fell into the patient.The metal flake was immediately removed from the patient and the area was irrigated.The procedure was completed successfully.No report of injury.
|
| |
|
Additional Manufacturer Narrative
|
|
V.Mueller has requested the device subject of the reported event be returned for evaluation.The snowden-pencer diamond-tip wire/pin cutter instructions for use states, "inspection/maintenance: inspect devices before each use for broken, cracked, tarnished surfaces, movement of hinges, and chipped or worn parts.If any of these conditions appear, do not use the device.Return devices to an authorized repair service center for repair or replacement." a 3-year complaint review indicates this to be an isolated event.A follow-up mdr will be submitted when additional information becomes available.
|
| |
|
Additional Manufacturer Narrative
|
|
The snowden-pencer diamond-tip wire/pin cutter was returned to the supplier, xka - kapp surgical gmbh for evaluation.The evaluation indicated that three chippings were noted on the cutting edge of the device.A specific root cause could not be determined.No additional issues have been reported.
|
| |
|