| Catalog Number |
TRM-310 |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
09/05/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Healthcare professional reported a right side "exchange from textured to smooth due to concern with the product." additionally, healthcare professional reported a capsular contracture baker grade unknown.Device has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture baker grade unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the device analysis grid, the assessments of the complaint are: ¿ capsular contracture: unable to observe as it is not related to the device.¿ anxiety-product/procedure: unable to observe as it is not related to the device.Additional observations: ¿ yellow biological tissue observed on the surface of the shell.¿ observed cloudy on the gel.¿ observed deformation on the device.No further actions are required since no manufacturing issue is observed.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported a capsular contracture baker grade iii.
|
| |
|