• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. TI-DBLE LEAD CORT 5.0X70MM SCR; ROD, FIXATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. TI-DBLE LEAD CORT 5.0X70MM SCR; ROD, FIXATION Back to Search Results
Catalog Number 14-405070
Medical Device Problem Code Unsealed Device Packaging (1444)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/27/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
It was reported that when opening the sterile package, it was noticed that the inner packaging seal was open.No adverse event has been reported as a result of the malfunction.Attempts to obtain additional information have been made; however, no more is available at this time.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Additional Manufacturer Narrative
Limited investigation: this complaint cannot be confirmed as the product was found to be conforming.Review of the work instructions found the pouch, is not validated to be sealed.No product failure was found as the product is within specifications.No further investigation or action is required after review.If any further information which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TI-DBLE LEAD CORT 5.0X70MM SCR
Common Device Name
ROD, FIXATION
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key17959713
Report Number0001825034-2023-02304
Device Sequence Number15214869
Product Code HSB
UDI-Device Identifier00887868092718
UDI-Public(01)00887868092718(17)2505311(10)020210
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
K063570
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2025
Device Catalogue Number14-405070
Device Lot Number020210
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/29/2023
Supplement Date Received by Manufacturer11/14/2023
Initial Report FDA Received Date10/18/2023
Supplement Report FDA Received Date12/01/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/12/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1