• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. PS TIBIAL ARTICULAR SURFACE PROVISIONAL; KNEE INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. PS TIBIAL ARTICULAR SURFACE PROVISIONAL; KNEE INSTRUMENT Back to Search Results
Model Number N/A
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the device was fractured.Attempts to obtain additional information have been made; however, no more is available.
 
Additional Manufacturer Narrative
(b)(4) h3: the complaint device has been returned, but the device investigation has not yet been completed.Once the evaluation is completed, a supplemental medwatch will be submitted.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.H6: mechanical g4-provisional top.Complaint sample was evaluated and the reported event was confirmed.Visual examination of the returned product identified sign of use and fracture.Device history record (dhr) was reviewed and no discrepancies relevant to the reported event were found.The summary of investigations for complaints associated with the tibial articular surface provisional (tasp) fracture.Fractured tasps were analyzed by optical microscopy revealing hackle marks and river lines in the case of bending overload; and hackle marks, river lines and striations were found in the case of low cycle fatigue culminating in a bending overload failure type.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PS TIBIAL ARTICULAR SURFACE PROVISIONAL
Common Device Name
KNEE INSTRUMENT
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key17959977
Report Number0001822565-2023-02897
Device Sequence Number7119630
Product Code JWH
UDI-Device Identifier00889024238824
UDI-Public(01)00889024238824(10)63523165
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K113369
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 03/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number42517400710
Device Lot Number63523165
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/02/2023
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/27/2023
Supplement Date Received by Manufacturer02/28/2024
Initial Report FDA Received Date10/18/2023
Supplement Report FDA Received Date03/05/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/10/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient SexUnknown
-
-