| Model Number |
100/897/080 |
| Medical Device Problem Code |
Inflation Problem (1310)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/20/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Other text: investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that the balloon would not inflate.They tried several times, but nothing worked.No medical intervention was required.
|
| |
|
Additional Manufacturer Narrative
|
|
No sample was returned for investigation, a photo of the reported condition was added for investigation.According to what is observed in the photo, it is not possible to confirm if the product has any leaks or any condition that would cause the cuff to fail to inflate.The complaint was not confirmed.The complaint report offered insufficient details to determine whether this product functioned as intended or was used in a manner consistent with its instructions for use (ifu) or failed to meet product specifications.Lacking any additional evidence, this complaint has been closed as unconfirmed.If the product is returned, icu medical will reopen this complaint for further investigation.A device history (dhr) review was performed subsequent to the manufacturing of the device and prior to its release.No problems or issues were identified during this dhr review.
|
| |
|
Additional Manufacturer Narrative
|
|
D4: udi updated.Udi related data quality updates only.
|
| |
|