| Model Number |
JP-2214 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
09/14/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
This complaint was forwarded to the manufacturing facility where it is currently still under investigation.A follow up report will be filed once the results have been completed.
|
| |
|
Event or Problem Description
|
|
The drain tube broke when the surgeon was removing it from the patient.It left about a 5" piece inside the patient requiring patient to be returned to surgery for removal.The patient was returned to surgery for removal and administered medication for anesthesia and recovery.He was discharged in stable condition to skilled nursing.
|
| |
|
Additional Manufacturer Narrative
|
|
No sample or photo were provided for evaluation therefore the supplier was unable to determine root cause for the reported issue.The batch record for lot number p22114035 was manufactured on 4/24/2023 with an expiration date of 4/24/2028.In-process controls contains tensile strength test of every batch of the drains.The batch passed tensile strength test (tear test) in the connection area.All the results were higher than minimum requirement.Cardinal health will continue to monitor and trend all similar reported product related issues.
|
| |
|