| Model Number |
430035-11 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/26/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a da vinci-assisted surgical procedure, the single-port monopolar curved scissors (mcs) tip fell off the single-port mcs instrument.There was no report of fragment(s) falling inside the patient.A backup instrument tip was used.The procedure was completed with no report of patient injury.Intuitive surgical, inc.(isi) followed up with the customer and obtained the following additional information: the instrument and single-port mcs tip were inspected prior to use, and no damage was found.The customer confirmed that there was no holes/tears/missing material on the mcs tip.They also did not notice any damage/missing piece on the mcs instrument.The mcs tip would not be returned.There was no report of fragments or mcs tip falling from the instrument while used within a patient.There was no report of injury to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation is in progress to determine the cause of this reported event.An rma was issued to evaluate the intuitive surgical, inc.(isi) device.Additional information is being gathered to determine the contribution of the device to the customer reported issue.A follow-up mdr will be submitted if the tip cover is returned (post failure analysis evaluation) or if additional information is received.A review of the provided image was performed by an intuitive surgical, inc.(isi) failure analysis engineer.The following additional information was provided: from the images provided, there does not appear to be any obvious material missing.The images provided appear to exhibit damage to the retaining cover, above where the internal retaining bayonet feature is.This damage is typically due to misinstallation of the tip accessory, a few possible causes: twisting the retaining cover before the accessory is properly seated on the instrument.An under installed sheath could lead to this, as the tip accessory cannot fully seat.Twisting the retaining cover too much and forcing the retaining bayonet out of the channel on the mcs instrument.Twisting the retaining cover when the accessory is not properly aligned (black stripe on the accessory¿s cover and the white stripe on the instrument need to align).
|
| |
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) received a da vinci product to perform failure analysis.The single port (sp) monopolar curved scissors (mcs) tip was analyzed and found to have the retaining tip cover cracked.The mcs tip was installed on an in-house mcs instrument and it was not recognized by the system.The issue persisted after reinstalling the mcs tip.During installation of the mcs tip to the in-house instrument, it was found that the retaining cover of the instrument tip could not be locked or secured.Visual inspection showed signs of cracking around one of the locking tabs of the accessory tip cover.No missing material was observed.The complaint was confirmed by failure analysis.
|
| |
|
Event or Problem Description
|
|
Refer to h10/h11 for follow-up information.
|
| |
|
Event or Problem Description
|
|
Refer to h10/h11 for follow-up information.
|
| |
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) completed further evaluation on the sp monopolar curved scissors (mcs) tip involved with this complaint.Advanced failure analysis confirmed the initial failure analysis (fa).The tip accessory was installed onto an in-house instrument and it was found that the disposable retaining cover was able to freely rotate on the instrument's instrument tube adapter.Because the cover is able to freely rotate, the cover is unable to securely engage with the instrument tube adapter.When the instrument's grip knob was manually actuated with the tip accessory installed, the accessory was pushed off of the instrument and failed the fit test.As mentioned in the initial fa , when the accessory is installed onto an in-house instrument, the instrument is unable to pass accessory detection on the system due to the accessory being unable to securely install onto the instrument.Due to this, the complaint of "during surgery, the tip cover fell off" cannot be fully replicated, but is plausible.To clarify, the metal scissor portion of the accessory should be retained assuming the scissor portion was correctly installed onto the instrument's ball socket and the instrument is not damaged.Visual inspection of the tip accessory's disposable retaining cover found one of the retaining features, also known as a lance, to be missing.The piece was not returned with the accessory and measures approximately 0.0735" x 0.0425".Additionally, the grey portion of the retaining cover surrounding the second retaining feature was found to be cracked, which allowed for the retaining feature to move outwards.The second retaining feature was not missing, but since it is able to move outwards due to the cracking, it is unable to securely engage with the instrument's instrument tube adapter.The root cause for the broken retaining cover and cracked retaining cover are both attributed to the user, and is typically caused by mis-installation of the tip accessory.This includes attempting to rotate the retaining cover to lock the cover when it is not fully seated, or over-rotating the retaining cover.Potential reasons for improper seating could be due to an instrument sheath not installed all the way, or by forcing the tip accessory on when the inner scissor portion or the coupling have not yet engaged.
|
| |
|
Search Alerts/Recalls
|