| Catalog Number |
254511455 |
| Medical Device Problem Code |
Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/11/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the non-cem fb tib drill bushing hole to small won't accept the drill.The event did not happen in surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary : an analysis of the product could not be performed since a physical sample was not received for evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.However, if the product is received at a later date, the investigation will be updated as applicable.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).His report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information that has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint (b)(4).Investigation summary: it was reported, that the non-cem fb tib drill bushing, hole to small.Won't accept the drill.The event did not happen in surgery.The device associated with this report was returned to depuy synthes for evaluation.Visual inspection of the returned sample revealed, that the att fb tib drill bushing por, was found a light stripped for the edge of the circumference.This condition cause, that the mating device cannot be slide for all the hole.Therefore, the reported condition can be confirmed.The observed, condition was identified, as an end of life indicator.Damage consistent with repeated use and servicing.The damage found on the edge of the circumference could be, due to the drill functioning when removing or inserting it.This can potentially cause damage in the hole.Properly handling and attention to the approved use of the device diminishes the risk of failure.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use.The device must be properly inspected, prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions and inspection procedures.A dimensional inspection was performed, for the att fb tib drill bushing por and met specifications.A functional test was performed, with the ping gage 9.35 mm, and it slid correctly through the inner circle of the device.However, it could not cross the outlet, due to the damage observed, at the hole's exit.The overall complaint was confirmed.As the observed, condition of the att fb tib drill bushing por, would have contributed to the complained issue.Based on the investigation findings, the potential cause is traced to end of life.And it has been determined, that no corrective and/or preventative action is required.As part of depuy synthes quality process, all devices are manufactured, inspected and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|