• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION JETSTREAM SC; CATHETER, PERIPHERAL, ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION JETSTREAM SC; CATHETER, PERIPHERAL, ATHERECTOMY Back to Search Results
Model Number 45006
Medical Device Problem Codes Entrapment of Device (1212); Mechanical Problem (1384); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/26/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
E1 - initial reporter address 1: (b)(6).
 
Event or Problem Description
It was reported that the jetstream catheter and wire became stuck.A 1.85mm jetstream sc atherectomy catheter was selected for endovascular therapy the superficial femoral artery using contralateral approach.When advancing the catheter, the angle of the common iliac artery branch was tight and difficult to rotate the 7f destination to the contralateral side, and it was also noted there was trouble with balloon anchors.During the jetstream procedure, the rotation stopped when it was attempted to break through the stenosis where calcification was the most severe.There was no problem at the beginning of the lesion, but a lot of resistance was felt when the catheter was brought to the lesion after about 3 seconds.The jetstream also became stuck with the thruway wire, and the products had to be removed together.The procedure was completed using an another jetstream device.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JETSTREAM SC
Common Device Name
CATHETER, PERIPHERAL, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
model farm road
cork T12 Y K88
EI   T12 YK88
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key17966281
Report Number2124215-2023-58028
Device Sequence Number7754328
Product Code MCW
UDI-Device Identifier08714729889861
UDI-Public08714729889861
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K133023
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2025
Device Model Number45006
Device Catalogue Number45006
Device Lot Number0032133381
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/26/2023
Initial Report FDA Received Date10/19/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/01/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-