| Catalog Number |
82096-01 |
| Medical Device Problem Code |
Difficult to Remove (1528)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/27/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a mildly calcified, mildly tortuous internal carotid artery that is 70% stenosed.The 10tx40x136 implanted xact stent was noted to be entangled with an unspecified guide wire, making it impossible to recover the embolic protection system (eps).A catheter was used to assist to removed the eps/ guide wire.The xact stent remained successfully implanted.There was no adverse patient effects reported and no clinically significant delay in the procedure.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.The investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely that after successful implantation of the xact stent and removal of the xact catheter interaction with the unspecified guide wire inadvertently resulted in the reported difficult to remove.There is no indication of a product quality issue with respect to manufacture, design or labeling.
|
| |
|