| Model Number |
SYM12F |
| Medical Device Problem Codes |
Loss of or Failure to Bond (1068); Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Codes |
Foreign Body In Patient (2687); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/28/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
According to the reporter, during a laparoscopic intraperitoneal onlay mesh abdominal incisional hernia surgery, the first two firings from the tacker were performed normally.From the third to the sixth firing, there was penetration through the mesh but fixation could not be performed normally.This resulted to damage from the mesh.The ones that penetrated the abdominal wall and penetrated the mesh were retained inside the body because there was a high possibility of bleeding if tacks were forcibly removed.From the seventh firing onwards, mesh could be fixated normally so the mesh was still kept inside the patient.Since the issue led to insufficient tacks, a competitor device was used to resolve the issue.
|
| |
|
Additional Manufacturer Narrative
|
|
D10 concomitant product: abstack15 abs fixation device 15 tacks (lot#: n3b0617y).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a laparoscopic intraperitoneal onlay mesh abdominal incisional hernia surgery, the first two firings from the tacker were performed normally.From the third to the sixth firing, there was penetration through the mesh but fixation could not be performed normally.This resulted to damage to the mesh.The ones that penetrated the abdominal wall and penetrated the mesh were retained inside the body because there was a high possibility of bleeding if tacks were forcibly removed.From the seventh firing onwards, mesh could be fixated normally so the mesh was still kept inside the patient.Since the issue led to insufficient tacks, a competitor device was used to resolve the issue.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
New information has been received pertaining to the event.This event has been reassessed and the reportability has been determined to be a malfunction.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|
|
|