MAUDE Adverse Event Report: ZIMMER BIOMET, INC. JGRKNT SHORT RIGID SZ2 IMP/DRL; FASTNER, FIXATION, NONDEGRADABLE, SOFT TISSUE/SPORTS MEDICINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. JGRKNT SHORT RIGID SZ2 IMP/DRL; FASTNER, FIXATION, NONDEGRADABLE, SOFT TISSUE/SPORTS MEDICINE Back to Search Results
Catalog Number 110005315
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 05/08/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that while the surgeon was pulling on the sutures of the implant the sutures fractured.The anchor remained implanted in the patient.The surgeon used three additional implant and the sutures fractured on all three and the anchors remained implanted in the patient.There was no harm reported to the patient or plan to explant the anchors from the patient.
 
Additional Manufacturer Narrative
(b)(4) multiple mdr reports were filed for this event, please see associated reports: 0001825034-2023-02450 0001825034-2023-02452 d10: medical products: item#: 110005315, jgrknt short rigid sz2 imp/drl; lot#: unknown item#: 110005315, jgrknt short rigid sz2 imp/drl; lot#: 0002402381 g2: foreign: australia no product was returned, or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history record(s) identified no deviations or anomalies during manufacturing.Medical records were not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JGRKNT SHORT RIGID SZ2 IMP/DRL
Common Device Name
FASTNER, FIXATION, NONDEGRADABLE, SOFT TISSUE/SPORTS MEDICINE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key17967321
Report Number0001825034-2023-02451
Device Sequence Number15094363
Product Code MBI
UDI-Device Identifier00880304568204
UDI-Public(01)00880304568204(17)270214(10)0002402496
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K110145
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number110005315
Device Lot Number0002402496
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/27/2023
Initial Report FDA Received Date10/19/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/14/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10 NARRATIVE.
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-