BOSTON SCIENTIFIC CORPORATION MUSTANG; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number 24674
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/08/2023
Type of Reportable Event Malfunction
Event or Problem Description
Reportable based on device analysis completed on 28-sep-2023.It was reported that balloon leak occurred.The 80% stenosed target lesion was located in the mildly tortuous and mildly calcified radial artery.A 3.0 x 40, 75cm mustang balloon catheter was advanced for dilatation.However, during the second inflation at 14 atmospheres, it was noted that the pressure pump could not fully inflate the balloon.After withdrawal, the balloon was inflated again but there was liquid leakage flowing from the surface of the balloon.The procedure was completed with a different device.There were no complications reported and the patient status was stable.However, returned device analysis revealed a balloon pinhole.
 
Additional Manufacturer Narrative
A2 - age at time of event: 18 years or older.(b)(6).Device evaluated by mfr: the device was received for analysis.A visual examination identified that the balloon was not folded which indicates that the device was subjected to positive pressure.Blood was identified inside the balloon which is evidence of a device leak.The returned device was attached to an encore inflation unit.Positive pressure was applied when liquid was observed to be leaking from a balloon pinhole located approximately 18mm proximal of the distal markerband.The rated burst pressure for this device is 24 atmospheres.A visual examination observed no issues or damage to the tip of the device.A visual and tactile examination found no kinks or damage the shaft of the device.A visual examination observed no issues with the markerbands of the device.Both markerbands were undamaged and present on the shaft of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MUSTANG
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key17970081
Report Number2124215-2023-58212
Device Sequence Number7754469
Product Code FGE
UDI-Device Identifier08714729793946
UDI-Public08714729793946
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K103751
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/04/2023
Device Model Number24674
Device Catalogue Number24674
Device Lot Number0026298403
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/25/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/28/2023
Initial Report FDA Received Date10/19/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/04/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
Patient Weight48 KG
-
-