| Catalog Number |
ASC4251-01 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/28/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference case-(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during set up, the starvac 90¿s tip detached.There was a s+n back up device available.There was a non-significant surgical delay and there was no patient involved.
|
| |
|
Additional Manufacturer Narrative
|
|
H10 h3, h6: the reported device was received for evaluation.Visual inspection observed the returned instrument shows no manufacturing abnormalities.The electrode has been used and fractured.The fractured pieces were returned in a specimen jar.There are signs of a sharp instrument touching or grasping the wand.Bio debris is present.Product was out of the original packaging.No packaging returned.A functional evaluation found the device displayed settings (1,7) when plugged in.Plasma and coagulation were generated on the remaining portions of the electrode.The resistance measured at 0.90 kilo ohms.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause has been associated with unintended use of the device.Factors that could have contributed to the reported event include: (1) hitting the device tip against a hard surface.(2) using the device as a lever to enlarge surgical site.(3) mechanical displacement of tissue through applied force.No containment or corrective actions are recommended at this time.
|
| |
|
Search Alerts/Recalls
|