CARDINAL HEALTH GSTRO FEED TBE W/Y PRT 20FR; TUBES, GASTROINTESTINAL (AND ACCESSORIES) Back to Search Results
Model Number 8884720205
Medical Device Problem Code Unintended Deflation (4061)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/27/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An investigation is currently underway.Upon completion, the results will be forwarded.
 
Event or Problem Description
The customer reported that during the endoscopy procedure they noticed that the retention balloon was punctured.There was no patient harm reported.Additional information was received from the customer and stated that during the procedure, it was noticed that the balloon had a malfunction, the balloon found pierced and leaked water.The customer further clarified that the balloon was not punctured by the operator during the procedure.
 
Additional Manufacturer Narrative
The lot number was provided, and the device history record (dhr) was reviewed indicating that the product was released accomplishing all quality standards.The product was manufactured on july 2, 2021.A sample was not received for the investigation.Because a sample was not returned, we were unable to perform a follow-up investigation to include functional and visual evaluations to confirm the issue and root cause analysis.However, as a part of continuous improvements, a supplier corrective action request (scar) has been issued to the supplier to address the reported condition through a more robust investigation; results of the investigation will be documented through the referred capa.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GSTRO FEED TBE W/Y PRT 20FR
Common Device Name
TUBES, GASTROINTESTINAL (AND ACCESSORIES)
Manufacturer (Section D)
CARDINAL HEALTH
777 west street
mansfield MA 02048
Manufacturer (Section G)
CARDINAL HEALTH
calle 9 sur no. 125 cuidad ind
tijuana 22500
MX   22500
Manufacturer Contact
jill saraiva
777 west street
mansfield, MA 02048
5086183640
MDR Report Key17970094
Report Number9612030-2023-03839
Device Sequence Number7754893
Product Code KNT
UDI-Device Identifier10884521007291
UDI-Public10884521007291
Combination Product (Y/N)N
Initial Reporter CountryBR
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number8884720205
Device Catalogue Number8884720205
Device Lot Number2117351664
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/18/2023
Supplement Date Received by Manufacturer10/18/2023
Initial Report FDA Received Date10/19/2023
Supplement Report FDA Received Date11/02/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/02/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-