| Catalog Number |
SSX-580 |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
07/10/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Healthcare professional reported a left side capsular contracture, baker grade iii.Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.The event of "capsular contracture, baker grade iii" is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture, baker grade iii.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the device analysis grid, the assessments of the complaint are: ¿ capsular contracture : unable to observe since it is a medical event and is not related to the device.Additional observations: ¿ crease observed on the device.¿ deformation observed on the device.No further actions are required as no issues with the manufacturing process are observed.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported a left side capsular contracture baker grade iii.Device has been explanted and not replaced.
|
| |
|