| Catalog Number |
6021-0435 |
| Medical Device Problem Codes |
Material Erosion (1214); Detachment of Device or Device Component (2907); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Codes |
Ambulation or Postural Difficulties (2544); Insufficient Information (4580)
|
| Date of Event |
09/27/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient's right hip was revised due to significant trunnion damage and partial dissociation of the femoral head.The stem and head were revised.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient's right hip was revised due to significant trunnion damage and partial dissociation of the femoral head.The stem and head were revised.
|
| |
|
Additional Manufacturer Narrative
|
|
Reported event: an event regarding disassociation involving accolade stem that was mated with a lfit v40 cocr head was reported.The event was confirmed.Method & results: device evaluation and results: visual inspection of the returned devices indicated that damage consistent with the loss of taper lock was observed on the head and stem.Clinician review: no medical records were received for review with a clinical consultant device history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the reported lot.Conclusion: it was reported that the patient was revised due to disassociation.Visual inspection of the returned devices indicated damage consistent with the loss of taper lock on the head and stem.The reported accolade stem was mated with a lfit v40 cocr head.The cocr head has been identified to be within scope of nc and capa.No further investigation for this event is required at this time.If additional information become available to indicate further evaluation is warranted, this record will be reopened.
|
| |
|