• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS COAGUCHEK ® XS SYSTEM; PROTHROMBIN TIME MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS COAGUCHEK ® XS SYSTEM; PROTHROMBIN TIME MONITOR Back to Search Results
Catalog Number 03666794001
Medical Device Problem Codes Display Difficult to Read (1181); No Display/Image (1183); Activation Problem (4042)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/25/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
On the call, a display check couldn't be done on the meter due to loss of power.The meter was provided for investigation.The investigation is ongoing.
 
Event or Problem Description
We received an allegation about a display issue on a coaguchek xs meter.The patient alleged that the meter turned on with a faded display too faint to read then it turned off.He replaced the batteries with new ones.The patient wasn't able to run a test as the meter did not power back on.
 
Additional Manufacturer Narrative
The device code (a code) was updated.The patient's device was provided for investigation.The device could not be turned on.The device data and the fault memory could not be read out.The circuit board was tested for damage or contamination.The battery contacts and the printed circuit board were contaminated by the liquid from the leaked battery which had penetrated and corroded the solder contacts.The root cause was contamination of the contacts due to improper handling or maintenance.The investigation did not identify a product problem.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK ® XS SYSTEM
Common Device Name
PROTHROMBIN TIME MONITOR
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key17976556
Report Number1823260-2023-03372
Device Sequence Number2454208
Product Code GJS
UDI-Device Identifier04015630945689
UDI-Public04015630945689
Combination Product (Y/N)Y
Initial Reporter StateWV
Initial Reporter CountryUS
PMA/510(K) Number
K062925
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 11/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue Number03666794001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/17/2023
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/25/2023
Supplement Date Received by Manufacturer11/02/2023
Initial Report FDA Received Date10/20/2023
Supplement Report FDA Received Date11/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexMale
-
-