| Catalog Number |
03666794001 |
| Medical Device Problem Codes |
Display Difficult to Read (1181); No Display/Image (1183); Activation Problem (4042)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/25/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
On the call, a display check couldn't be done on the meter due to loss of power.The meter was provided for investigation.The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
We received an allegation about a display issue on a coaguchek xs meter.The patient alleged that the meter turned on with a faded display too faint to read then it turned off.He replaced the batteries with new ones.The patient wasn't able to run a test as the meter did not power back on.
|
| |
|
Additional Manufacturer Narrative
|
|
The device code (a code) was updated.The patient's device was provided for investigation.The device could not be turned on.The device data and the fault memory could not be read out.The circuit board was tested for damage or contamination.The battery contacts and the printed circuit board were contaminated by the liquid from the leaked battery which had penetrated and corroded the solder contacts.The root cause was contamination of the contacts due to improper handling or maintenance.The investigation did not identify a product problem.
|
| |
|