| Model Number |
BIO-COMP SWIVELOCK SP, 5.5X24.5MM |
| Medical Device Problem Code |
Defective Device (2588)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/21/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a surgery the anchor couldn`t be fixed, it was rotating.According to the surgeon no harm for patient, operator or third party occurred.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6 complaint is confirmed.Visual inspection identified that the bio-screw at the distal end of the tip suture anchor, bio-composite swivelock® sp, 5.5 x 24.5 mm, self-punching , vented had broken.The eyelet and driver returned no issues were found.Functional testing noted that the driver works smoothly as required.The method used to prepare the bone, as well as the bone quality encountered, was not provided.Per dfu-0087-eo revision 2.G.Precautions.3.Make sure to use the recommended drill bit or punch to create a bone socket.8.Self-punching pushlock and swivelock suture anchors only: ensure that the angle of anchor insertion is perpendicular to the bone.The most likely cause for the reported failure can be attributed to improper bone preparation, misalignment insertion of the device during insertion; prying/leveraging the device during insertion.
|
| |
|
Search Alerts/Recalls
|