• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. UNIVERSAL GLENOID - CENTRAL SCREW 20MM; SHOULDR PROSTH, REVERSE CONFIG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. UNIVERSAL GLENOID - CENTRAL SCREW 20MM; SHOULDR PROSTH, REVERSE CONFIG Back to Search Results
Model Number UNIVERSAL GLENOID - CENTRAL SCREW 20MM
Medical Device Problem Codes Biocompatibility (2886); Patient Device Interaction Problem (4001)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 09/19/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event or Problem Description
It was reported that after an initial surgery on (b)(6) 2023 the patient developed a severe infection which affected the patient for 6 months.A revision surgery was performed on (b)(6) 2023 to remove the involved implants.No further information received.
 
Additional Manufacturer Narrative
Complaint confirmed.One unpackaged ar-9165-20 universal glenoid - central screw 20mm serial/batch number 20.03580 was received for investigation.Visual evaluation found that the returned device's hex geometry was damaged, maybe due to excessive tightening forces.The method of bone preparation employed during the procedure, as well as the bone quality encountered, was not provided.A cause for the reported failure may be a patient-specific event.Per dfu-0189-9 at revision 0.C.Contraindications.3.Foreign body sensitivity.Where material sensitivity is suspected, appropriate tests should be made, and sensitivity ruled out prior to implantation.4.Any active infection or blood supply limitations.D.Adverse effects.1.Infections, both deep and superficial.2.Foreign body reactions.7.Allergies and other reactions to device materials.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIVERSAL GLENOID - CENTRAL SCREW 20MM
Common Device Name
SHOULDR PROSTH, REVERSE CONFIG
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
emily shafer
8009337001
MDR Report Key17976814
Report Number1220246-2023-08282
Device Sequence Number2453894
Product Code PHX
UDI-Device Identifier00888867057937
UDI-Public00888867057937
Combination Product (Y/N)N
Initial Reporter CountrySW
PMA/510(K) Number
K142863
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 12/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/31/2025
Device Model NumberUNIVERSAL GLENOID - CENTRAL SCREW 20MM
Device Catalogue NumberAR-9165-20
Device Lot Number20.03580
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/25/2023
Supplement Date Received by Manufacturer09/25/2023
Initial Report FDA Received Date10/20/2023
Supplement Report FDA Received Date12/19/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-