| Model Number |
UNIVERS REVERS SUTURE CUP, 42 (+2 RIGHT) |
| Medical Device Problem Codes |
Biocompatibility (2886); Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
Unspecified Infection (1930)
|
| Date of Event |
09/19/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that after an initial surgery on (b)(6) 2023 the patient developed a severe infection which affected the patient for 6 months.A revision surgery was performed on (b)(6) 2023 to remove the involved implants.No further information received.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint confirmed.One unpackaged ar-9502f-42rcpc univers revers suture cup, 42 (+2 right) serial/batch number (b)(6) was received for investigation.A bullet nose screw was attached to the cup.Visual evaluation of the suture cup noted no issues with the returned device, only signs of wear; evaluation of the bullet nose screw found damaged in the hex head and thread, causing geometry loss.The method of bone preparation employed during the procedure, as well as the bone quality encountered, was not provided.A cause for the reported failure may be a patient-specific event.Per dfu-0189-9 at revision 0.C.Contraindications.3.Foreign body sensitivity.Where material sensitivity is suspected, appropriate tests should be made, and sensitivity ruled out prior to implantation.D.Adverse effects.1.Infections, both deep and superficial.2.Foreign body reactions.7.Allergies and other reactions to device materials.
|
| |
|