| Catalog Number |
09180443001 |
| Medical Device Problem Code |
Application Program Problem (2880)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/09/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
The initial reporter stated there was an issue with the cobas infinity software.The customer alleged that patient results observed in the evolution (validation), patient history (validation), and history/evolution charts (order search) sections of the cobas infinity software are not presented in a chronologically correct order.The software uses the order creation date instead of the sample collection date to place samples in chronological order.The customer alleges this could possibly give a misleading picture of the patient's progress to the clinicians reviewing the sample results through the evolution, patient history, or history sections of the software.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer states that the same issue is occurring in other sections of the cobas infinity software.The 'last result in previous orders' column in work areas\validation also follows the same faulty logic.The 'value of previous result' field used in variables/rule engine also follows the same faulty logic.The investigation confirmed the alleged behavior as a software issue since parameters in the cobas infinity software are ordered based on the "creation date", causing interpretation issues.
|
| |
|