BOSTON SCIENTIFIC CORPORATION ENDOTAK RELIANCE G; IMPLANTABLE LEAD Back to Search Results
Model Number 0262
Medical Device Problem Code Failure to Convert Rhythm (1540)
Health Effect - Clinical Code Electric Shock (2554)
Date of Event 10/06/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the health care professional (hcp) requested technical services (ts) review of this implantable cardioverter defibrillator (icd) device data to confirm device operation.The hcp reported that multiple episodes of bursts of anti-tachycardia pacing (atp) and shocks were delivered appropriately due to ventricular arrythmia.The shocks and atps therapies were unable to convert the arrythmia, with an episode that recorded this icd device to have delivered all its programmed therapy.It was also noted that during a defibrillation threshold (dft) test performed had difficulties converting the rhythm.The icd and lead remain in service.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOTAK RELIANCE G
Common Device Name
IMPLANTABLE LEAD
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646 -260
*   00646-2602
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key17980689
Report Number2124215-2023-58520
Device Sequence Number11385750
Product Code LWS
UDI-Device Identifier00802526601828
UDI-Public00802526601828
Combination Product (Y/N)Y
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P910073/S077
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date12/17/2021
Device Model Number0262
Device Catalogue Number0262
Device Lot Number100243
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/06/2023
Initial Report FDA Received Date10/20/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/17/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Other;
Patient Age50 YR
Patient SexUnknown
-
-