• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SURGICAL SPECIALTIES MEXICO S. DE. R.L. DE C.V. STRATAFIX; SH 2-0 MONODERM 20CM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SURGICAL SPECIALTIES MEXICO S. DE. R.L. DE C.V. STRATAFIX; SH 2-0 MONODERM 20CM Back to Search Results
Model Number SXMD1B405
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 09/20/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A batch review of the finished good lot and components confirmed there were no quality issues noted throughout the incoming inspection, manufacturing, in-process, or final inspection.The needle component is supplied by ethicon.A review of the broken needle will be performed by a 3rd party laboratory.No samples, or photographs of the broken device were provided for review.No third party evaluation report was received to date.If samples, photos or the report becomes available at a later time, they will be reviewed, and the results will be included in the file.A follow-up report will be submitted at that time.The bending, fracturing, breaking of a needle can occur when needles are gripped with a needle holder, forceps or some type of surgical instrument on or near the swaged area or near the tip of the device, when excessive force is applied, when the device(s) are used in applications involving tortuous tissue or with a needle tip design that may not be appropriate for the specific tissue or procedure.The stress on the attachment section is greatly reduced when the needle is held in an area one-third (1/3) to one-half (1/2) of the distance from the swaged end to the point as indicated in the ifu for this product.Without reviewing the actual broken needle, testing sterile devices from the same finished good lot, receiving the results of the third party evaluation of the needle component or receiving details regarding the preoperative preparation of the device, tools utilized to grasp the needle component or tools utilized in conjunction with the medical device (robotics), the size trocar used compared to the size needle on the device, procedure performed or the surgeon¿s technique, a definitive root cause cannot be determined at this time.
 
Event or Problem Description
The health authority reported to our distributor the needle broke when the surgeon attempted to sew the peritoneum.Changed to another one to complete the surgery.No additional information could be provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STRATAFIX
Common Device Name
SH 2-0 MONODERM 20CM
Manufacturer (Section D)
SURGICAL SPECIALTIES MEXICO S. DE. R.L. DE C.V.
corredor tijuana rosarito 2000
24702 b, ejido francisco villa
tijuana 22235
MX  22235
Manufacturer (Section G)
SURGICAL SPECIALTIES CORPORATION
247 station drive
suite ne1
westwood MA 02090
Manufacturer Contact
ronald giannangelo
corredor tijuana-rosarito 2000 #24702b
ejido francisco villa
tijuana 22235
MX   22235
MDR Report Key17981300
Report Number3010692967-2023-00050
Device Sequence Number3518362
Product Code GAM
UDI-Device Identifier10705031228061
UDI-Public(01)10705031228061(20)12(17)240831(10)AAFA204
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 10/20/2023
Report Date (Section F) 10/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2024
Device Model NumberSXMD1B405
Device Lot NumberAAFA204
Was the Report Sent to FDA? No
Distributor Facility Aware Date09/20/2023
Device Age4 YR
Event Location Hospital
Date Report to Manufacturer10/12/2023
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/12/2023
Initial Report FDA Received Date10/20/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/25/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient SexUnknown
-
-