According to the reporter, during a laparotomy pancreatoduodenectomy, during resection of pancreatic head lesion, during use of the first stapler, the surgeon felt that the device did not squeeze sufficiently and the device did not fire.The second device was squeezed firmly but the staples did not properly form the b-shape and it partially fired.A manual stapler was used for additional resection at the unstapled site and there was an extended incision less than one inch due to the position of the tumor resection where it was difficult to allow margins.A manual stapler ensured that there was a cutting allowance hence resulting on expanded incision.
|
|
D10: concomitant product: ta6048s - ta6048s ta 60-4.8 reloadable stapler, lot# p2m0504.H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the loading unit had been partially applied.This can be replicated by an incomplete cycle of the instrument.It was reported that the device did not fire and did not provide the expected tactile feedback.The reported issues were confirmed.The product analysis noted evidence that the device was not used as intended.This issue may occur when the firing stroke is not completed during application.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: to fire the instrument, squeeze the handle a second time until it reaches a solid stop and it locks in the back position.If the "instrument" handle is partially squeezed during firing then released, bleeding may occur as the instrument was not fully fired.The black release button may be used to open the instrument at any time during approximation.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|