| Catalog Number |
786426 |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/25/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that the customer found the ureteral stent defective during the procedure.Per follow up information received on 05oct2023, according to the operating theatre, pulling the cord also broke the stent.The event occurred before the stent was inserted into the patient.A 2nd stent has since been broken in a similar manner, a stent with the same lot number.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event is confirmed cause unknown.Visual evaluation noted received 1 photo sample that shows top view of stent being separated into two separate pieces.Therefore, product does not meet specifications.Although an exact root cause could not be determined a potential root cause could be user does not handle with care, bends sharply.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "avoid improper handling of stent such as bending, kinking, tearing, etc.Misuse could damage the overall integrity of the stent." "care should be exercised when removing the stent from inner polybag so as not to cause tearing or fragmentation." "the insertion of a ureteral stent should only be done by those individuals who have comprehensive training in the techniques and risks of the procedure." section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that the customer found the ureteral stent defective during the procedure.Per follow up information received on 05oct2023, according to the operating theatre, pulling the cord also broke the stent.The event occurred before the stent was inserted into the patient.A 2nd stent has since been broken in a similar manner, a stent with the same lot number.
|
| |
|