C.R. BARD, INC. (COVINGTON) -1018233 BARD® INLAY OPTIMA® URETERAL STENT WITH NICORE¿ GUIDEWIRE Back to Search Results
Catalog Number 786426
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/25/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that the customer found the ureteral stent defective during the procedure.Per follow up information received on 05oct2023, according to the operating theatre, pulling the cord also broke the stent.The event occurred before the stent was inserted into the patient.A 2nd stent has since been broken in a similar manner, a stent with the same lot number.
 
Additional Manufacturer Narrative
The reported event is confirmed cause unknown.Visual evaluation noted received 1 photo sample that shows top view of stent being separated into two separate pieces.Therefore, product does not meet specifications.Although an exact root cause could not be determined a potential root cause could be user does not handle with care, bends sharply.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "avoid improper handling of stent such as bending, kinking, tearing, etc.Misuse could damage the overall integrity of the stent." "care should be exercised when removing the stent from inner polybag so as not to cause tearing or fragmentation." "the insertion of a ureteral stent should only be done by those individuals who have comprehensive training in the techniques and risks of the procedure." section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that the customer found the ureteral stent defective during the procedure.Per follow up information received on 05oct2023, according to the operating theatre, pulling the cord also broke the stent.The event occurred before the stent was inserted into the patient.A 2nd stent has since been broken in a similar manner, a stent with the same lot number.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® INLAY OPTIMA® URETERAL STENT WITH NICORE¿ GUIDEWIRE
Common Device Name
URETERAL STENT
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key17981586
Report Number1018233-2023-07623
Device Sequence Number14070156
Product Code FAD
UDI-Device Identifier10801741015042
UDI-Public(01)10801741015042
Combination Product (Y/N)N
Initial Reporter CountryBE
PMA/510(K) Number
K043193
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number786426
Device Lot NumberNGGX5612
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/05/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/05/2023
Supplement Date Received by Manufacturer01/29/2024
Initial Report FDA Received Date10/20/2023
Supplement Report FDA Received Date01/30/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/23/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-