| Model Number |
CLH-250 |
| Medical Device Problem Codes |
Failure to Power Up (1476); Defective Component (2292)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/25/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for evaluation and the customer's allegation was confirmed.The device evaluation found another reportable malfunction reported in the b5 section.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
The customer reported to olympus that during reprocessing, the halogen light source does not start.There was no procedure involved and no reports of patient or user harm.The device was returned for evaluation.During the device evaluation, another reportable malfunction was found: fan failure.This medical device report (mdr) is being submitted to capture both reportable malfunctions: one reported by the customer and the other found during evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.Since the device is more than 15 years old, the device history record was unable to be reviewed.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Based on the results of the investigation, the root cause of the event was unable to be identified.The event likely occurred due to a defective lamp socket and defective fan.Olympus will continue to monitor field performance for this device.
|
| |
|