Catalog Number 04.045.034S
Medical Device Problem Code No Apparent Adverse Event (3189)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 01/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: b3: only the event year is known.D2b: additional device product codes: hsb, jds.H3, h6: the device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on an unknown date, the physician reported that a short rfna which he had implanted on (b)(6) 2022 had broken at the junction between the most distal screw of the proximal interlocking screws.The surgeon removed the broken hardware on (b)(6) 2022 and revised the construct with a nail and plate.Upon product return to the manufacturer and examination on october 11, 2023, it was identified that two screws were stripped.This report involves one locking screw for im nail ø 5mm/ 34mm/ xl25/ sterile.This is report 3 of 3 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.A manufacturing record evaluation was performed: product code: 04.045.034s-us, lot: 656p864: it was electronically reviewed and no nonconformances / manufacturing irregularities were identified during the manufacturing process.The product was released on: february 23, 2022.Manufacturing site: (b)(6).Expiry date: january 31, 2032.A product investigation was completed: visual analysis of the returned sample revealed that the locking screw was found the middle threads stripped.No other defect was found.A dimensional inspection was not performed as it is not applicable to the complaint condition.The current and manufactured to drawing was reviewed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOCKING SCREW FOR IM NAIL Ø 5MM/ 34MM/ XL25/ STERILE
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key17984191
Report Number8030965-2023-13306
Device Sequence Number20875824
Product Code HWC
UDI-Device Identifier10886982294395
UDI-Public(01)10886982294395
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K201346
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number04.045.034S
Device Lot Number656P864
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/11/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/11/2023
Supplement Date Received by Manufacturer11/16/2023
Initial Report FDA Received Date10/23/2023
Supplement Report FDA Received Date11/21/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/04/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
LOCKSCR F/NAILS Ø5 L32 XL25; LOCKSCR F/NAILS Ø5 L64 XL25; LOCKSCR F/NAILS Ø5 L64 XL25; LOCKSCR F/NAILS Ø5 L68 XL25; RFNA/ 9MM / 160MM/ STANDARD BEND/ STER; UNK - PLATES: TRAUMA
Patient Age62 YR
Patient SexFemale
-
-