| Model Number |
98DS014 |
| Medical Device Problem Code |
Air/Gas in Device (4062)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/27/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Conclusion not yet available, evaluation in process.A follow-up report will be submitted as soon as the investigation is complete.Oscor inc.Is submitting the above report to comply with 21 cfr part 803, the medical device reporting regulation.The report is based upon information obtained by oscor inc.Which the company may not have been able to investigate or verify prior to the date the report was required by fda.This report does not constitute an admission that the device, oscor inc., or its employees, caused or contributed to the event described in this report.Nor does this report reflect a conclusion by fda, oscor inc., or its employees, that the report constitutes an admission that the device, oscor inc., or its employees caused or contributed to the event described in this report.
|
| |
|
Event or Problem Description
|
|
The complaint affects the loader, while flushing with the device in it, there were air bubbles entering the loader from the hemostatic valve.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: b5 full description of event reported the loader from a 14f occlutech delivery sheath set was returned from the customer.The sheath and dilator were not returned for analysis.According to the event description summary, air bubbles were entering the loader from the hemostatic valve.Upon evaluation of the returned loader, it was found that the hemostatic valve looked normal as per the drawing.The loader was manually leak tested with the syringe connected to the stopcock and the tip occluded.The loader did not exhibit leakage when tested using this method.The loader was then tested using the iso 11070 for liquid leakage test method.The device was occluded at the distal tip and pressurized to 38kpa and held at this pressure for 30 seconds and no drops of liquid were observed from the hemostatic valve.The loader was further pressurized beyond 300kpa and no leakage was observed from the stopcock or sideport tubing.Per qa procedure (breezeway ii loader in-process and final inspection) sampling plan: inspect per ansi z 1.4, gen level 1, normal, and aql 0.40 perform leak test according to procedure.Per ifu: instructions for use when delivery sheath is used with loader: a loader is to be used at a physician's discretion to open the valve of a delivery sheath for ease of insertion of a diagnostic and/or therapeutic device into the delivery sheath.· follow directions for use from section 8.1 for general use lines 1-6.· completely flush the loader assembly via the side port with three-way stopcock with either saline or heparinized saline to ensure it is free of air prior to use to avoid air embolism to the patient.· inspect loader for air bubbles.Constantly flush loader to prevent air bubbles while loading a diagnostic and/or therapeutic device.· before connecting the loader containing the device to the delivery sheath, remove the dilator and guidewire from the delivery sheath.· secure the loader containing the device using the screw in connector on the delivery sheath hub.Figure 3 and figure 4.Caution: the lock ring on the loader must be screwed tightly against the delivery sheath.This prevents detachment of the loader from the delivery sheath when pushing the device through.Returned device analysis revealed the loader was within manufacturing specifications.The loader did not leak from the hemostatic valve or anywhere else when tested manually.The loader also met the iso leakage test method requirement with no leakage found anywhere (including the hemostatic valve).No manufacturing rejects or anomalies of this type were recorded in the device history record.The introducer sheath and loader passed all in-process and qa final inspection steps before shipping to the customer, including visual, dimensional, mechanical and leak testing.The reason for return cannot be confirmed.No manufacturing defects were found.The samples are dispositioned per the applicable customer agreement as per procedure pro-001019.In conclusion, based on the information available at this time it cannot be confirmed that the product failed to meet requirements.Oscor inc.Is submitting the above report to comply with 21 cfr part 803, the medical device reporting regulation.The report is based upon information obtained by oscor inc.Which the company may not have been able to investigate or verify prior to the date the report was required by fda.This report does not constitute an admission that the device, oscor inc., or its employees, caused or contributed to the event described in this report.Nor does this report reflect a conclusion by fda, oscor inc., or its employees, that the report constitutes an admission that the device, oscor inc., or its employees caused or contributed to the event described in this report.
|
| |
|
Event or Problem Description
|
|
The customer reported the event occurred during patient contact.The complaint affects the loader, while flushing with the device in it, there were air bubbles entering the loader from the hemostatic valve.After flushing the device, air got in the loader again while bringing the loader to the sheath.As soon as the pusher slightly moved, there was air entering the loader again from the hemostatic valve.The physician holds the pusher and loader completely still so that it worked.The procedure was completed successfully.
|
| |
|