| Catalog Number |
10-210 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
09/22/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left implant ruptured.Device remains implanted.
|
| |
|
Event or Problem Description
|
|
The device has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Device analysis: ased on the device analysis grid, the assessments of the complaint are: ¿ rupture: observed broken on posterior side assessed as fold flaw opening and missing piece of shell assessed as inconclusive.Additional observations: ¿ observed stress marks on the device.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left implant ruptured.The device has been explanted and replaced.
|
| |
|