MENTOR TEXAS MENTOR SILTEX ROUND MODERATE PROFILE; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
|
Back to Search Results |
|
| Catalog Number |
3542670 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Deformity/ Disfigurement (2360)
|
| Date of Event |
08/31/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The mentor failure analysis lab has received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.A manufacturing record evaluation is in progress.Once completed, a supplemental report will be submitted.Reason for device explant and/or reoperation: deflation.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a female who underwent a breast augmentation revision with a mentor siltex round moderate profile, 425cc, saline breast implant experienced bilateral deflation postoperatively.The issue was diagnosed through physical examinations.As a result, patient underwent bilateral replacements with cat: 3501650 on (b)(6) 2023.This report relates to the left side.Note: mentor received two deflated devices, therefore both implants are reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation completed on october 31, 2023: the product was returned to mentor for evaluation.Mentor then conducted visual inspection and microscopic examination of the returned device.Visual analysis of the returned sample determined that the sal siltex rnd diap pkg 425cc breast implant was found to have (2) two tears.Tear (a) extending from the posterior to the anterior view measuring approximately 2.0 cm and tear (b) on the anterior view measuring approximately 1.2 cm.Microscopic examination was performed on the edges of both tears, and parallel striations were found in the whole area of the tears.Parallel striations are consistent with markings made by a sharp object perforating the implant shell.Based on the information currently available, microscopic examination of the returned product indicates that the implant could have been damaged during or subsequent to implantation.The product insert data sheet cautions to not allow cautery devices or sharp instruments, such as scalpels, suture needles, hypodermic needles, hemostats, adson forceps, or scissors to contact the device during the implantation or other surgical procedures.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Capa activity was not required.The manufacturing record evaluation (mre) was reviewed, and no anomalies were found related to this complaint.In addition, the mre review verifies that the device was manufactured in accordance with documented specification and procedures.Manufacturer¿s reference number (b)(4).
|
| |
|