| Model Number |
518-019 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Cardiac Perforation (2513); Pericardial Effusion (3271)
|
| Date of Event |
09/26/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
D4): device lot number, expiration date unk.Partial udi populated.H3): the device was discarded, thus no investigation could be completed.H4): device manufacture date unk because lot number unk.H6): cardiac perforation is a known risk of complication with use of the lld.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
|
| |
|
Event or Problem Description
|
|
A lead extraction procedure commenced to remove a right atrial (ra) and a right ventricular (rv) lead due to bacteremia.Spectranetics lead locking devices (lld #2) were inserted into each lead to provide traction.Beginning with the rv lead, a spectranetics 11f tightrail rotating dilator sheath was used, advancing to the tip of the lead, and the lead came free.Approximately 2 minutes later, the patient''s blood pressure dropped, a pericardial effusion was detected in the region of the rv apex, and an rv perforation was suspected.Rescue efforts began, including pericardiocentesis, and 250 cc of blood was drained from the area, stabilizing the patient with no further intervention required.The procedure continued, extracting the ra lead successfully, and the patient survived the procedure.This report captures the lld #2 providing traction to the rv lead when a suspected perforation occurred, requiring intervention.There was no alleged malfunction of any spectranetics devices in use during the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Udi related data quality updates only.D4): correction: removing primary di number from udi# field.The lot number was not provided, thus an entire udi# is unavailable.G3): on 11jul2024, the manufacturer received fda email correspondence identifying a discrepancy in the udi information (section d of the initial mdr).Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
|
| |
|
Search Alerts/Recalls
|
|
|