| Catalog Number |
30660105S |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Perforation (2001)
|
| Date of Event |
09/27/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Based on the available information the device will not be returned therefore an evaluation of the device cannot be performed.Should additional information become available, it will be provided in a supplemental report.H3 other text : device disposition is unknown.
|
| |
|
Event or Problem Description
|
|
As reported: "after surgery, the u-lag screw cut out the femoral head with u-blade deformation and the revision surgery was performed.".
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event could be confirmed, with the provided radiographs.From the received images we can clearly see the deformation on the u-blade one of the blades is bent completely and we can see the impact on the complete blade as we see the material being folded.This impact could have been caused by excessive force application while the blade is stuck or hitting a hard surface.Further, the device is required for the complete investigation.With the available radiographs, a formal medical opinion was sought: the event can be confirmed but with limited information provided, further evaluation could not be conducted.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.If the device is returned or if any additional information is provided, the investigation will be reassessed.
|
| |
|
Event or Problem Description
|
|
As reported: "after surgery, the u-lag screw cut out the femoral head with u-blade deformation and the revision surgery was performed.".
|
| |
|