• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CORDIS US CORP. SUPER TORQUE MB; CATHETER, INTRAVASCULAR, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CORDIS US CORP. SUPER TORQUE MB; CATHETER, INTRAVASCULAR, DIAGNOSTIC Back to Search Results
Catalog Number 532598C
Medical Device Problem Code Structural Problem (2506)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/20/2023
Type of Reportable Event Malfunction
Event or Problem Description
As reported, the marker band slid along the 5f 110cm 8 side holes (sh) super torque pigtail marker band (mb) angiographic catheter during the intervention.The marker band did not fall off the device during use in the patient; it dislocated from the catheter and slid prior to measurement during use in the patient.All the marker bands were present on the catheter when removed from the patient.Another super torque was used to complete the procedure.There was no reported patient injury.The intended procedure was not provided.Access site and vessel characteristics at target site were not provided.The device was stored and handled per the instructions for use (ifu).The device was not exposed to a hot climate or room temperature.There were no visible signs of device/packaging damage prior to use or anomalies noted when the device was taken out of the package.No excessive force was applied to the device during withdrawal from the package.The event did not cause a clinically relevant increase in the duration of the procedure nor a condition that required hospitalization or significant prolongation of existing hospitalization.The device was discarded at site.A procedural cd or radiological report is not available for review.
 
Additional Manufacturer Narrative
Additional information is pending and will be submitted within 30 days upon receipt.
 
Additional Manufacturer Narrative
As reported, the marker band slid along the 5f 110cm 8 side holes (sh) super torque pigtail marker band (mb) angiographic catheter during the intervention.The marker band did not fall off the device during use in the patient; it dislocated from the catheter and slid prior to measurement while inside the patient.All the marker bands were present on the catheter when removed from the patient.Another super torque was used to complete the procedure.There was no reported patient injury.The intended procedure was not provided.Access site and vessel characteristics at target site were not provided.The device was stored and handled per the instructions for use (ifu).The device was not exposed to a hot climate or room temperature.There were no visible signs of device/packaging damage prior to use or anomalies noted when the device was taken out of the package.No excessive force was applied to the device during withdrawal from the package.The event did not cause a clinically relevant increase in the duration of the procedure nor a condition that required hospitalization or significant prolongation of existing hospitalization.Without the return of the device or images for analysis, the reported customer event ¿marker band (super torque mb-offset/out of position-prior to/during measurement¿ could not be confirmed.Excessive manipulation of the catheter while the device was trapped in the patient¿s vasculature may have contributed to the reported event.Users are trained to inspect for signs of damage prior to and during use.Any product with damage is not to be used.Information for safety is provided in the products labeling with the intent to make the user aware of the risks.According to the instructions for use (ifu), although not intended as a mitigation of risk, ¿contraindications: do not use the super torque mb catheter in procedures where entrapment of the catheter between endovascular devices and the vessel wall may occur, for example endovascular aortic repair (evar) procedures.Manipulation of the catheter under excessive friction due to interaction with other devices or while trapped in the vasculature, can lead to stretching or elongation of the catheter.Stretching or elongation of the catheter during endovascular procedures could result in the marker bands moving along the catheter.In extreme cases, marker bands may come off the catheter and dislodge into the vascular system.¿ without the return of the device, the root cause could not be conclusively determined.However, an investigation to address marker band movement has already been initiated.No further action will be taken at this time.
 
Event or Problem Description
The device was discarded at site.A procedural cd or radiological report is not available for review.
 
Additional Manufacturer Narrative
After further review, section d.4 primary unique device identification (udi) number has been corrected accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPER TORQUE MB
Common Device Name
CATHETER, INTRAVASCULAR, DIAGNOSTIC
Manufacturer (Section D)
CORDIS US CORP.
14201 nw 60th avenue
miami lakes, florida 33014
Manufacturer (Section G)
CORDIS US CORP.
14201 nw 60 avenue
miami lakes, florida 33014
Manufacturer Contact
karla castro
santiago troncoso 808
juarez, chihuahua 
7863138372
MDR Report Key17995363
Report Number9616099-2023-06616
Device Sequence Number14938883
Product Code DQO
UDI-Device Identifier10705032012041
UDI-Public(01)10705032012041(17)250228(10)18092733
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K915836
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2025
Device Catalogue Number532598C
Device Lot Number18092733
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/03/2023
Supplement Date Received by Manufacturer10/31/2023
08/14/2024
Initial Report FDA Received Date10/24/2023
Supplement Report FDA Received Date11/01/2023
08/14/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/09/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-