| Model Number |
7927 |
| Medical Device Problem Codes |
Use of Device Problem (1670); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Codes |
Abrasion (1689); Bruise/Contusion (1754); Fall (1848); Skin Tears (2516)
|
| Date of Event |
10/13/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Patient was prone on the table, secured with some kind of disposable strap, which broke.Patient fell off table, experiencing a contusion.
|
| |
|
Additional Manufacturer Narrative
|
|
Following the information obtained from the investigation, no problems with the device were found.Lateral tilt function was engaged instead of intended reverse trendelenburg function that cause the table to tilt unexpectedly.A disposable safety strap was used that is not manufactured by mizuho osi and that did not hold the patient during the unexpected table tilt thus causing the patient fall.Minor skin tears and abrasion were observed in the patient following the fall.The patient was treated immediately and given discharge the same day.
|
| |
|
Event or Problem Description
|
|
Patient was prone on the table, secured with some kind of disposable strap, which broke.Patient fell off table, experiencing a contusion.
|
| |
|