• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: APPLIED MEDICAL RESOURCES CB030, 5MMX35CM EPIX LAPA SCISSOR, 10/BX; LAPAROSCOPE, GYNECOLOGIC (AND ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

APPLIED MEDICAL RESOURCES CB030, 5MMX35CM EPIX LAPA SCISSOR, 10/BX; LAPAROSCOPE, GYNECOLOGIC (AND ACCESSORIES) Back to Search Results
Model Number CB030
Medical Device Problem Codes Tear, Rip or Hole in Device Packaging (2385); Device Contaminated During Manufacture or Shipping (2969)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/28/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The event device is anticipated to return.A follow-up report will be provided upon completion of the investigation.
 
Event or Problem Description
Name of procedure being performed: na (before use).Detailed description of event: consumable was contaminated as there was a break of sterility by another consumable that was placed on top of it and the sterile field.Consumable not used on patient.No clinical impact to the patient as a result of the event.Open another pack to resolve the situation.Type of intervention: na (before use).Patient status: na (before use).
 
Event or Problem Description
Name of procedure being performed: na (before use).Detailed description of event: consumable was contaminated as there was a break of sterility by another consumable that was placed on top of it and the sterile field.Consumable not used on patient.No clinical impact to the patient as a result of the event.Open another pack to resolve the situation.Type of intervention: open another pack.Patient status: no clinical impact to the patient.
 
Additional Manufacturer Narrative
The event unit was returned to applied medical for evaluation with no inner packaging.As the inner packaging was not returned, the complainant¿s experience of a sterility break is unable to be confirmed.Applied medical is unable to determine if the inner packaging exhibited any non-conformances that could have contributed to the reported event.In the absence of the inner packaging, it is difficult to determine if the reported event was caused by a manufacturing non-conformance or circumstantial factors at the time of use.Applied medical has reviewed the details surrounding the event and is unable to determine the exact root cause of the event.The probability and criticality of harm resulting from this failure have been evaluated and were found to be at an acceptable level.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CB030, 5MMX35CM EPIX LAPA SCISSOR, 10/BX
Common Device Name
LAPAROSCOPE, GYNECOLOGIC (AND ACCESSORIES)
Manufacturer (Section D)
APPLIED MEDICAL RESOURCES
22872 avenida empresa
rancho santa margarita CA 92688
Manufacturer Contact
farah azmi
22872 avenida empresa
rancho santa margarita, CA 92688
9497138710
MDR Report Key17996353
Report Number2027111-2023-00662
Device Sequence Number12224075
Product Code HET
UDI-Device Identifier00607915110123
UDI-Public(01)00607915110123(17)260310(30)01(10)1483234
Combination Product (Y/N)N
Initial Reporter CountrySN
PMA/510(K) Number
K062169
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 12/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/10/2026
Device Model NumberCB030
Device Catalogue Number101419701
Device Lot Number1483234
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/28/2023
Supplement Date Received by Manufacturer09/28/2023
Initial Report FDA Received Date10/24/2023
Supplement Report FDA Received Date12/01/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/11/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-