Catalog Number 252-13-06
Medical Device Problem Codes Naturally Worn (2988); Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Pain (1994); Synovitis (2094); Insufficient Information (4580); Joint effusion (4849)
Date of Event 08/31/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Pending investigation.
 
Event or Problem Description
As reported via legal documentation the patient had a right knee replacement on (b)(6) 2015.Approximately 2 years and 9 months after the initial procedure the patient had a right knee revision on (b)(6) 2018.There is no other patient demographic or medical history available.There is no information on the surgical procedure or patient outcome.There is no device return.There are no photos or other images of the device provided.No additional information is available.The patient has filed a short-form complaint in a multidistrict litigation titled in re: exactech polyethylene orthopedic products liability litigation, mdl no.3044 (ngg) (mmh), and pending in the eastern district of new york.Per the transfer order creating this multidistrict litigation, ¿plaintiffs¿allege that their knee or hip replacement devices¿failed prematurely because of degradation of the device¿s polyethylene component, which resulted in the premature removal (or planned removal) of the prosthesis at issue.¿ because the patient has filed a suit in this multidistrict litigation, the patient appears to allege that the patient was injured as a result of premature wear of an exactech polyethylene device.
 
Additional Manufacturer Narrative
D4: added catalog number, expiration date, serial number, and udi.H6: corrected health effect - clinical code, medical device problem code, component code, type of investigation, investigation findings, and investigation conclusions.Manufacturer narrative updated: the reason for the revision cannot be confirmed from the information provided but may be the result of prosthesis wear or due to inclusion of the polyethylene in the packaging recall.However, the reported prosthesis wear could not be confirmed and potential contributions of manufacturing, user, or patient-related considerations to the event could not be assessed as the devices were not available for evaluation and no images, radiographs, or relevant clinical information was provided.
 
Additional Manufacturer Narrative
A2, a3: added date of birth and gender.H6: added impact code 4629.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated: g3/g4, h6 the following sections were corrected: d1/d2a/d2b, h6 mdr section codes updated/corrected: b, e the reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear and femoral loosening or due to inclusion of the polyethylene in the packaging recall.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images and radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNI ALL POL TIB COMP RT MEDIAL 8MM SZ 6
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH INC.
2320 nw 66th ct.
gainesville FL 32653
Manufacturer Contact
kate jacobson
MDR Report Key17996660
Report Number1038671-2023-02593
Device Sequence Number12233339
Product Code JWH
UDI-Device Identifier10885862065865
UDI-Public10885862065865
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K040889
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2018
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 01/17/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/15/2020
Device Catalogue Number252-13-06
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/14/2023
Supplement Date Received by Manufacturer09/13/2024
Not provided
01/14/2025
Initial Report FDA Received Date10/24/2023
Supplement Report FDA Received Date09/16/2024
11/21/2024
01/17/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age83 YR
Patient SexMale
-
-