Catalog Number 337146
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/29/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H.3.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that while using bd facs¿ lyse wash assistant tere were erroneous results acquired.The following was reported by the initial reporter: samples do not lyse properly and remain red.The problem occurs in all samples in a rack.
 
Additional Manufacturer Narrative
The following fields have been updated with corrected and/or additional information: reporting office: becton dickinson and company bd biosciences.Reporting office contact: (b)(6).Manufacturing site contact: (b)(6).H.6: based on the investigation results, the reported issue for the incomplete lysing was confirmed.Confirmed >.Investigation results that were performed on the indicated failure mode were: the potential cause was determined to be defective valve unit and clogged cell wash.The device history record was reviewed to confirm the instrument met all the manufacturing specifications prior to release.The field service representative performed a series of works including replacement of the valve unit and cell wash assembly.After the replacement, a series of tests and verifications were performed, and the instrument was confirmed to be operating as per specifications.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for identification of emerging trends.
 
Event or Problem Description
It was reported that while using bd facs¿ lyse wash assistant there were erroneous results acquired.The following was reported by the initial reporter: samples do not lyse properly and remain red.The problem occurs in all samples in a rack.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD FACS¿ LYSE WASH ASSISTANT
Common Device Name
STATION, PIPETTING DILUTING CLINICAL USE
Manufacturer (Section D)
BECTON DICKINSON AND COMPANY BD BIOSCIENCES
2350 qume drive
san jose CA 95131
Manufacturer (Section G)
BECTON, DICKINSON AND COMPANY, BD BIOSCIENCES
2350 qume drive
san jose CA 95131
Manufacturer Contact
fahmy razak - mdr
2350 qume drive
san jose, CA 95131
4089542435
MDR Report Key17996684
Report Number2916837-2023-00206
Device Sequence Number17600021
Product Code JQW
UDI-Device Identifier00382903371464
UDI-Public00382903371464
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number337146
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/29/2023
Supplement Date Received by Manufacturer03/22/2024
Initial Report FDA Received Date10/24/2023
Supplement Report FDA Received Date03/26/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/28/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-